Hanna Instruments (M) Sdn Bhd
11A, Jalan PJS 11/20,
Bandar Sunway,
46150 Petaling Jaya,
Selangor, Malaysia.
+603-5638 9940
Penang Branch Office
Hanna Instruments (M) Sdn Bhd
N0. 303-2-27, Krystal Point,
Jalan Sultan Azlan Shah,
11900 Sungai Nibong,
Penang, Malaysia.
+604-638 4558
+604-645 4558
penang@hannamalaysia.com

Sabah Branch Office
Hanna Instruments (M) Sdn Bhd

No. 4-1, 1st Floor,
Plaza Kingfisher,
Jalan Plaza Kingfisher 5,
Inanam, 88450 Kota Kinabalu,
Sabah, East Malaysia.
088-382 941
088-382 942
sabah@hannamalaysia.com


HI3001 Flow-thru Conductivity Probe

Previous 15 / 44 Next

HI3001 Flow-thru Conductivity Probe Probe Process Controllers
RM 1,207.00
500 g
In Stock


This four-ring flow-thru probe measures conductivity in industrial applications with a platinum sensor and NTC sensor for Automatic Temperature Compensation. With it's durable design, you can count on this electrode to stand up to your process environment.

  • Four-ring probe measures conductivity with platinum sensors.
  • Houses an NTC sensor for Automatic Temperature Compensation.
  • Built for everyday, demanding use in industrial applications.

Four-Ring Probe

  • The HI3001 is a potentiometric sensor, utilizing four-ring technology to achieve accurate EC measurements. The four-ring technology allows for a wide range of measurement with a single probe as compared to amperometric two-pole probes that have a limited range and can suffer from polarization effects. The platinum sensors are protected by a PEI cover that can easily be removed for quick maintenance.

PVDF Body

  • Polyvinylidene fluoride (PVDF) is a plastic that is resistant to most chemicals and solvents. It has high abrasion resistance, mechanical strength, and resistance to ultraviolet and nuclear radiation. PVDF is also resistant to fungal growth.

Internal Temperature Sensor

  • The integrated NTC temperature sensor of the HI3001 probe is crucial for accurate EC measurements. Since the temperature of a solution can drastically change the amount of conductance, obtaining a fast and stable temperature measurement allows for an accurately compensated EC reading.

Multiple Cable Lengths

  • The HI3001D models feature a DIN connector for use with Hanna’s HI99XX series of wall-mounted controllers, along with varying lengths of cable. HI3001D comes with a DIN connector and 3 meters (9.9’) of cable attached. HI3001D/5 comes with a DIN connector and 5 meters (16.4’) of cable attached. HI3001D/10 comes with a DIN connector and 10 meters (32.8’) of cable attached.

Threaded Connection

  • The ½” front NPT threaded connection of the HI3001 allows for flow-thru mounting, while the ¾” back threads allow for submersion or pipe mounting without the need for additional hardware.

PEI Protective Cover

  • The protective cover is made of PEI and can be removed for quick maintenance. These probes can withstand temperatures up to 80°C (176°F) and 6 bars (87 psi) of pressure.

Specifications

Body Material PEI and PVDF
EC/TDS Temperature Compensation Range automatic, 0 to 60°C with NTC sensor
Max. Pressure 6 bar (87 psi)
Overall Length 89 mm (3.5”)
Cable 3 m (9.9’), 5 m (16.4’) and 10 m (32.8’) based on selection
Connector Type N/A (HI3001), DIN (HI3001D models)

 









Please leave your enquiry here, we will reply as soon as possible.
Name*  
Company Name  
Product Interested  
Quantity  
Email*  
Contact No.*  
Attachment  
*only support gif, jpeg, jpg, png, pdf
Messages*  

You have 0 items in you cart. Would you like to checkout now?
0 items
Switch to Mobile Version
Subscribe Newsletter